WindSelect+

The AmbientAirMonitoringSystem

Standards and Methods

 

Cartridge preparation

In the laboratory, one plane filter and two or three PU-foam filters are inserted into the cartridge and

the sampling standard is spiked to the plane filter.

 

Sampling procedure

The cartridges are connected with quick connectors, the 90 mm laminar tubes are mounted to the cartridges. The sampling data are reset and the wind sensor is connected.

 

After the respective configuration, the measurement is started. The device samples 4 m3/h automatically, the cartridge is selected dependening on the configuration of the sectors and the calm limit.

After the sampling period, e.g. 4 weeks, the sampling values are checked and noted. The cartridges are unmounted, sealed with the cap and transported to the laboratory, In the laboratory, the cartridges are evaluated, e.g. using EN 1948.

  • The minimum flow sucked through each cartridge has to be  50 m3, which results in a detection limit of 2 pg/m3 for dioxins.
     
  • Using TS EN 1948-4, I-TEQ PCB can be evaluated. Evaluation of PCBs needs a minimum sampling volume of 200 m3 per cartridge.

 

The cartridges are rinsed after the procedure and prepared for the next use.

 

 

 