isokinetic withdrawal of the flue gas |
|
|
use of glass fibre filter for dust precipitation |
|
|
2 stage column of adsorbent material for adsorbtion of gaseous dioxins |
The main differences are the use of dilution air and the special designed filter cartridge, which
allows to work without condenser. The condensation of a liquid phase is avoides and the dioxins are collected in the
dry filter cartridge for the following analysis.
The improvement of long term sampling leads to very low blank values. A volume of 100 to 400 m3 is
sampled during every measurement . The increase of the sampled volume decreases the detection limit to 0,001 ng
I-TEQ/m3.
The ETV test program has shown that I-TEQ values of the DioxinMonitoringSystem® are
within <10% uncertainty compared to EPA method 23A.