Design:
|
for industrial use
|
Sampling Method
|
|
Standard reference:
|
dilution method in full accordance to EN1948-1:(2006), extended for long term sampling
|
|
Filter System:
|
3 stage dry solid filters, separating dust and gas fraction
|
|
Sampling Probes:
|
2 special heated titanium zero-pressure probes
(for very small gas channels also with only one probe available)
|
|
Sampling Intervall:
|
usually 4 weeks (2 hours to 6 weeks)
(to measure I-TEQPCDD/F)
|
|
Range of method:
|
6 hours 0,01 - 100 ng TEQ/m³
6 weeks 0,0001 - 10 ng TEQ/m³
(depending on laboratory parameters)
|
|
Isokinetic Control:
|
direct at each actual sampling position
|
|
Remote Control:
|
Internet-TCP/IP4
|
Operating Environment
|
|
Temperature:
|
-10 to +45 °C
|
|
Humidity:
|
10 - 90 % relative, not condensing
|
Physical Dimensions
|
|
Control Unit:
|
62/62/171 cm (w/d/h)
|
|
Outer Sampling Unit:
|
52/28/60 cm (w/d/h) - outside the stack
|
|
Inner Sampling Unit:
|
depending on length of probes - inside the stack
|
|
Heated Tube:
|
length up to 70 m
|
Gas Parameters
|
|
Velocity:
|
2 to 20 m/s (up to 35 m/s)
|
|
Temperature:
|
up to 250°C (up to 450°C) *)
|
|
Humidity:
|
up to 350 g/m³ *)
|
|
Dust Content:
|
0-20 mg/m³ (up to 150 mg/m3)
|
|
Corrosive Gases:
|
limited according to layout
|
|
|
*) due to interdependence cross influence, these parameters have to be
taken into consideration together
|
External Supply
|
|
Main Power:
|
400 V 16 A, 3 phases + neutral + ground
alternatively
|
|
|
230 V 16 A, 1 phase + neutral + ground
for a maximum length of the tube of 12 meters
|
|
Compressed Air:
|
typical consumption: 2-6 m³/h dependent on humidity
|
Other
|
|
Maintenance Interval:
|
typically 12 months (dependent on application)
|
|
Installation:
|
outdoor or indoor
|